L1 /   
5
CH3041
LOQ, IDL
Instrument Detection Limit (IDL)
• The minimum concentration which may be
discriminated from zero.
1) IDL = 3 s (calib. blank)  s =standard deviation
eg. 12.00 ± 0.01    s = 0.01  \ 3s = 0.03
2) IDL S/N > 2:1 (with lowest level of analyte)
Limit of quantitation (LOQ) = 10 s (blank)
This is the limit of analyte that can be detected on the instrument with confidence.
Method Detection Limit (MDL) : obtained using spiked samples, the lowest level which gives a LOQ is the MDL. It gives the lowest level of quantitation for the entire analytical procedure not just the instrument.